Company record
Greiner Bio-One North America, Inc.
0 FDA warning letters since 2023 and 2 recalls in the last 12 months.
Get an email the day FDA posts anything new about Greiner Bio-One North America, Inc..
Recalls
| Reported | Class | Firm | Product and reason |
|---|---|---|---|
| Oct 28 | Class II | Greiner Bio-One North America, Inc. Device |
Brand Name: VACUETTE Product Name: LH Lithium Heparin Separator Model/Catalog Number: 454008P Product Description: Greiner VACUETTE Blood Collection Tube with Lithium Heparin and Gel Separator (13 x 75 mm, 3.5ml) Lot B26043FM may include EDTA contamination which introduces potassium while chelating calcium in patient samples, leading to falsely elevated potassium (K+) and falsely decreased calcium (Ca++) results which have the potential to cause hyperkalemia. |
| Nov 26, 2025 | Class II | Greiner Bio-One North America, Inc. Device |
Brand Name: Vacuette Product Name: 8 ml CAT Serum Separator Clot Activator Model/Catalog Number: 455071P Product Description: Greiner VACUETTE Blood Collection Tube with Clot Activator and Gel Separator and Standard S… Some of the Vacuette¿ CAT Serum Separator Clot Activator tubes were manufactured without gel separator, which could yield erroneous assay results and require the patient to return to the lab for recollection or repeat testing. This may result in delayed diagnosis, cause a delay in clinical decision-making or treatment and in some rare cases, yield incorrect results from less stable parameters. |